Skip to main content
. Author manuscript; available in PMC: 2015 Jan 27.
Published in final edited form as: Circulation. 2001 Jul 17;104(3):310–316. doi: 10.1161/01.cir.104.3.310

TABLE 2.

Frequency of Differences Beyond Clinically Acceptable Limits of Disagreement

Measurement Clinically Acceptable Limits Frequency Beyond Limits
Clinically* Acceptable Limits Frequency Beyond Limits
n % n %
Fractional shortening, % (n=724) Within ±2.5% 411 56.8 ±10% of mean 324 44.8
End-diastolic dimension, cm (n=724) Within ±0.15 cm 187 25.8 ±10% of mean   71   9.8
End-diastolic posterior wall thickness, cm (n=735) Within ±0.20 cm   44   6.0 ±10% of mean 479 65.2
End-diastolic septal thickness, cm (n=726) Within ±0.20 cm 118 16.3 ±10% of mean 504 69.4
*

Secondary standard.