Published in final edited form as: Curr Protoc Mol Biol. 2015 Jan 5;109:21.30.1–21.30.32. doi: 10.1002/0471142727.mb2130s109
Problem Potential cause Solution
Poor quality LC-MS run Problem with Sample or with HPLC column Run a peptide standard;
if standard = poor, replace HPLC
if standard = fine, problem with sample
Poor electrospray stability or overall low signal intensity Insufficient desalting More washes of sample during peptide recovery
Sudden increase in HPLC pressure Particulate debris in the sample Replace HPLC column, lines, junctions to autosampler, and/or the rotor and stator of the autosampler valve; Centrifuge peptide sample at 16000 x g, transfer top 90% of volume
Poor chromatography – fronting peaks (peak skews to earlier retention time) Column capacity overloaded due to overly concentrated peptide sample Dilute sample
Poor chromatography – tailing peaks (peak skews to later retention time) Residual reactive silanol groups in the stationary phase, or improper formulation of the mobile phases (pH) Replace HPLC column and/or HPLC buffers