Skip to main content
. Author manuscript; available in PMC: 2016 Jan 31.
Published in final edited form as: Pediatr Radiol. 2014 Jul 5;45(2):181–187. doi: 10.1007/s00247-014-3109-7

Table 1.

CT parameters. Lower CTDIvol is achieved in group B primarily by reductions in mAseff and kVp as compared to group A.

Traditional weight-based
protocols with FBP

(Group A)
Reduced-CTDIvol protocols
with FBP/IRT blend
(Group B)
Section thickness (mm) 3 3

Section interval (mm) 1.5 1.5

Rotation time (s) 0.5 – 0.8 0.5

Pitch 0.7 – 1.0 0.9

Reconstruction filter Standard B Standard B

Patient weight (kg) n mAseff kVp Mean
CTDIvol
(mGy)
n mAseff kVp Mean
CTDIvol
(mGy)

<9 3 70–75 120 4.6 4 56 100 2.2
9–17 3 80 120 4.9 1 52 100 2.1
5 3
18–26 3 90 120 5.5 5 58–60 100 2.3
7 0
27–35 2 100 120 6.3 2 64–68 100 2.5
4 7
36–44 1 120 120 7.3 2 75–80 100 3.1
9 7
45–67 5 150 120 9.3 6 95–96 120 6.3
8 1
68–81 1 200 120 12.2 1 119–120 120 7.9
6 2

FBP filtered back projection, IRT iterative reconstruction technique, mAseff effective milliampere-seconds, kVp peak kilovoltage, CTDIvol volume CT dose index, n number of subjects