Table 1.

Scan protocols for all phantom scans using the Philips Brilliance and the HD750 GE Discovery CT scanners. All protocols are helical scans.

Phantom Protocol Pitch mA Tube rotation time (s) Filter SFOV (cm) Slice thickness (cm) DFOV (cm) Recon kernel CTDIvol (mGy)
RMI phantom Philips 120kVp 0.938 375 0.75 B 60 3 36 B 19.9
GE 120kVp 0.969 450 0.5 Medium 50 2.5 36 Standard 21.2
GE GSI preset #35 0.969 630 0.5 Medium 50 2.5 36 Standard 20.2

Head phantom Philips 120kVp 0.688 202 0.75 B 60 3 25 B 26.2
GE 120kVp 0.516 280 0.6 Medium 50 2.5 25 Standard 26.0
GE GSI preset #32 0.516 375 0.6 Medium 50 2.5 25 Standard 26.0

Pelvic phantom Philips 120kVp 0.938 263 0.75 B 60 3.0 36 B 13.9
GE 120kVp 0.984 265 0.6 Medium 50 2.5 36 Standard 12.7
GE GSI preset #41 0.984 360 0.6 Medium 50 2.5 36 Standard 12.8

Spine phantom Philips 120kVp 0.938 300 0.75 B 60 3.0 50 B 15.9
GE 120kVp 0.969 265 0.7 Medium 50 2.5 50 Standard 17.1
GE GSI preset #25 0.969 375 0.7 Medium 50 2.5 50 Standard 17.0