Skip to main content
. 2015 Jan 29;9(1):014110. doi: 10.1063/1.4905874

TABLE I.

Details of five experiments with co-ordinated resistive pulse and optical measurements, including the number of identified events and their matching rate at different applied pressures. Numbers of events which show no corresponding events in the other trace are given in brackets. The Wilson score49 is a lower bound for the matching percentage at 97.5% confidence, assuming a binomial distribution.

Experiment Pressure (mm water) Number of events Wilson score (%)
TRPS Fluorescence Matching Double events
D1 10 ± 1 8 (0) 8 (0) 8 (100%) 0 76
U1 ∼−30 22 (0) 21 (0) 20 (91%) 1 78
D2 ∼70 58 (4) 54 (0) 54 (93%) 0 87
D3 100 ± 1 82 (3) 76 (1) 71 (87%) 4 82
D4 ∼150 83 (4) 71 (0) 63 (76%) 8 70