Comparison of the SAMBA semi-Q and Roche TaqMan v2 results for 488 clinical samples from London (n = 134), Malawi (n = 200), and Uganda (n = 154)a

Sample source and SAMBA Semi-Q result (copies/ml) No. with result with Roche TaqMan v2 (copies/ml)
<1,000 >1,000
London
    <1,000 95 1
    >1,000 2 36
Malawi
    <1,000 146 0
    >1,000 4 50
Uganda
    <1,000 91 5
    >1,000 3 55
a

Concordance between the two tests was 96.9% (473/488) overall, 97.8% (131/134) in London, 98.0% (196/200) in Malawi, and 94.8% (146/154) in Uganda.