. 2014 Nov;52(11):3878–3883. doi: 10.1128/JCM.02063-14

TABLE 2.

Sensitivities, specificities, and predictive values of the CAP/CTM with DBS or DPS for diagnosing ARV TFa,b

Specimen type No. of plasma DBS or DPS pairs Diagnostic threshold, copies/ml Sensitivity (CI) Specificity (CI) PPV (CI) NPV (CI)
DBS 823 >5,000 0.99 (0.96–1.00) 0.55 (0.51–0.59) 0.33 (0.29–0.37) 1.00 (0.98–1.00)
>1,000 0.99 (0.97–1.00) 0.26 (0.22–0.29) 0.29 (0.26–0.33) 0.99 (0.96–1.00)
DPS 546 >5,000 0.88 (0.81–0.94) 0.98 (0.96–0.99) 0.92 (0.85–0.97) 0.97 (0.95–0.98)
>1,000 0.91 (0.85–0.95) 0.96 (0.94–0.98) 0.89 (0.83–0.94) 0.97 (0.94–0.98)
a

Plasma VL is the reference standard for sensitivity, specificity, and predictive values of VL in DBS or DPS. Plasma, DBS, or DPS that had a VL that was undetectable or lower than the LOD were classified as negative (i.e., <5,000 copies/ml or <1,000 copies/ml) for ARV TF.

b

All confidence intervals (CIs) are reported at the 95% confidence level.