. 2014 Dec;52(12):4224–4229. doi: 10.1128/JCM.01553-14

TABLE 3.

Comparison between genomic loads measured by lytA rtPCR from sputum and from nasopharyngeal swabs

Performance No. of lytA rtPCR results of >10,000/ml/total no. (%) in sputum ofa:
No. of NP swab lytA rtPCR results of >8,000/ml/total no. (%) (n = 222)
Any quality (n = 222) Good quality (n = 42) Suboptimal quality (n = 180)
Sensitivity 63/81 (77.8) 22/28 (78.6) 41/53 (77.4) 64/81 (79.0)
Specificity 94/141 (66.7) 9/14 (64.3) 85/127 (66.9) 96/141 (68.1)
PPVb 63/110 (57.3) 22/27 (81.5) 41/83 (49.4) 64/109 (58.7)
NPV 94/112 (83.9) 9/15 (60.0) 85/97 (87.6) 96/113 (85.0)
Additional yield 47/141 (33.3) 5/14 (35.7) 42/127 (32.3) 45/141 (31.9)
a

Composite diagnostic standard: positive blood culture, induced-sputum culture or Gram stain, urine antigen, or whole-blood lytA rtPCR; for this comparison, the cutoff value of >8,000 copies/ml for NP swab lytA rtPCR criterion was not included in the composite diagnostic.

b

PPV, positive predictive value; NPV, negative predictive value; additional yield, 1−specificity (i.e., patients with sputum lytA rtPCR cutoff value of >10,000/ml or NPS lytA rtPCR cutoff value of >8,000/ml and a negative composite diagnostic standard represent false positives).