TABLE 2.

Laboratory performance levels of three rapid diagnostic tests for the dual detection of HIV and T. pallidum antibodies, compared to reference methodsa

RDT HIV antibody
T. pallidum antibody
% sensitivity (95% CI) % specificity (95% CI) % sensitivity (95% CI) % specificity (95% CI)
SD Bioline 97.9 (92.0–99.6) 100 (91.5–100) 93.0 (84.8–97.1) 100 (92.9–100)
ChemBio 98.9 (93.6–99.9) 98.1 (88.6–99.9) 95.3 (87.9–98.5) 100 (92.9–100)
MedMira 97.9 (92.0–99.6) 94.2 (83.1–98.5) 94.1 (86.3–97.8) 96.9 (88.2–99.5)
a

The reference HIV antibody method was the Siemens Advia Centaur HIV 1/O/2 enzyme immunoassay and the GS HIV-1 Western blot kit, and the T. pallidum antibody reference method was the Serodia TP-PA method.