TABLE 1.

Mean Artifact Scores of 2D Fast-Spin Echo (FSE) Images Versus Multiacquisition Variable-Resonance Image Combination Selective (MAVRIC SL) Images for 25 Matched Image Sets

Anatomy Imaged and
Imaging Plane
Mean Artifact Score (± SD)
p
2D FSE MAVRIC SL

Hip
 Coronal
  Reviewer 1 4.92 ± 0.28 2.77 ± 0.73 0.0002a
  Reviewer 2 5.00 ± 0.00 2.62 ± 0.51 0.0002a
 Axial
  Reviewer 1 4.92 ± 0.29 2.58 ± 0.79 0.0005a
  Reviewer 2 4.92 ± 0.29 2.67 ± 0.49 0.0005a
Femur
 Coronal
  Reviewer 1 4.69 ± 0.48 2.23 ± 0.60 0.0002a
  Reviewer 2 4.38 ± 0.51 1.85 ± 0.38 0.0002a
 Axial
  Reviewer 1 4.58 ± 0.67 2.92 ± 1.24 0.0029a
  Reviewer 2 4.58 ± 0.51 2.00 ± 0.00 0.0005a
Overall
 Hip
  Reviewer 1 4.92 ± 0.28 2.68 ± 0.75 < 0.0001a
  Reviewer 2 4.96 ± 0.20 2.64 ± 0.49 < 0.0001a
 Femur
  Reviewer 1 4.64 ± 0.57 2.56 ± 1.00 < 0.0001a
  Reviewer 2 4.48 ± 0.51 1.92 ± 0.28 < 0.0001a

Note—Artifact was scored by two reviewers on a 5-point scale from 1 (normal or barely visible artifact) to 5 (severe artifact and nonvisualization of bone-metal interface).

a

Statistically significant difference (p < 0.05).