Table 1.

Percentage improvements in the mean for speckle SNR, contrast and CNR between fundamental B-mode vs. SLSC and harmonic B-mode vs. HSCI across all three image quality groups.

Good
Quality
(n=28)
Medium
Quality
(n=34)
Poor
Quality
(n=29)
Overall
(n=91)
Speckle
SNR
Fundamental B-mode
vs. SLSC
62%* 46%* 41%* 50%*
Harmonic B-mode vs.
HSCI
109%* 73%* 48%* 72%*
Contrast Fundamental B-mode
vs. SLSC
10% 28%* 67%* 29%*
Harmonic B-mode vs.
HSCI
4.3% 20%* 79%* 17%*
CNR Fundamental B-mode
vs. SLSC
71%* 60%* 56%* 64%*
Harmonic B-mode vs.
HSCI
117%* 80%* 69%* 85%*
*

The signifies a statistically significant improvement with a p<0.01.