Skip to main content
. 2015 Jan 6;15(1):832–854. doi: 10.3390/s150100832

Table 1.

Parameters of the electric circuit model (mean + s.e.m): C is the double-layer capacitance, Rp the charge transfer resistance and Rs the path + saline resistance.

Electrode Diameter Electrode Area [mm2] Rs [kΩ] Rp [MΩ] C [nF]
100 μm 0.031 2.6 ± 0.1 83.5 ± 31.7 1.3 ± 0.1
300 μm 0.283 1.8 ± 0.02 4.8 ± 0.4 10.6 ± 0.4
500 μm 0.783 1.5 ± 0.02 1.3 ± 0.1 36.0 ± 3.1