Skip to main content
. Author manuscript; available in PMC: 2016 Apr 1.
Published in final edited form as: MAGMA. 2014 Aug 14;28(2):195–205. doi: 10.1007/s10334-014-0455-2

Fig. 4.

Fig. 4

a Schematic diagram of the 129-GHz microwave system used for DNP. The components within the dashed box (attenuator-3 port circulator-diode detector) can be bypassed during DNP. Insets photographs of the top section of the microwave system. b Relative 13C polarization level of trityl-doped [1-13C]pyruvic acid sample (see Table 1) as a function of microwave power at 4.6 T and 1.1 K where the threshold microwave power is around 20 mW. Data were plotted by obtaining the maximum 13C NMR intensity after microwave irradiation of the sample at the positive polarization peak. Microwave power was controlled by a variable attenuator and the power was measured with a diode detector. The dashed curve is a guide for the eyes