Skip to main content
. 2014 Apr 22;14(1):11. doi: 10.1186/1470-7330-14-11

Table 1.

Scan parameters of breast MRI study

  Technique TR/TE/FA Matrix FOV Thickness/Gap NEX SENSE Fat Sat. Scan time
DWI
2D-SE-EPI
5600/99/90
128 × 114(256R)
280 mm
4/0.4 mm
3
Factor 2.0
SPAIR
4 m 40 s
T2WI
2D-FSE
6200/80/90
256 × 198(256R)
280 mm
4/0.4 mm
2
Factor 2.0
SPAR
1 m 58 s
T1WI
2D-GRE
268/5.5/80
256 × 198(256R)
280 mm
4/0.4 mm
1
Factor2.0
Non
1 m 27 s
Dynamic
3D-Fast-GRE
7.7/3.8/12
352 × 334(512R)
280 mm
1.2/-0.6 mm
1
Factor2.0
SPAIR
70 s × 3
High-Resolution 3D-Fast-GRE 8.6/4.3/12 512 × 496(512R) 280 mm 1.21-0.6 mm 1 Factor2.0 SPAIR 3 mls