Skip to main content
. 2014 Dec;17(4):36–49. doi: 10.3831/KPI.2014.17.035

Table. 6. System suitability testing.

Parameter QUE BDMC DMC CUR
Mean RSD (%) Mean RSD (%) Mean RSD (%) Mean RSD (%)
Retention time, tR 3.970 0.021 13.840 0.027 15.230 0.025 16.723 0.021
Peak area 11221611 0.806 286851 0.654 1590498 0.651 5448675 0.711
Resolution, R - - 32.195 0.321 2.887 0.364 2.830 0.370
Tailing factor, Tf 1.369 0.108 1.501 0.261 1.165 0.144 1.081 0.051
Theoretical plate, N 8803.785 0.359 15552.398 0.865 13763.145 0.646 15568.252 0.910
Capacity factor, k’ 0.684 0.846 4.870 0.415 5.460 0.406 6.093 0.391

RSD, relative standard deviation; QUE, quercetin; BDMC, bisdemethoxycurcumin; DMC, demethoxycurcumin; CUR, curcumin. N, number of theoretical plates; k’, capacity factor; Mean of six replicate injections of quality control (QC) standard of 160 μg/mL.