Table 3. Sensitivity and Specificity of the Screening Devices for Detection of Strabismus and Amblyopia.
Characteristic | Resulta |
---|---|
PVS Test result, positive/negative | |
Strabismus, amblyopia, or both | 183/5 |
Neither | 14/97 |
Sensitivity | 0.97 (183/188) |
95% CI | 0.94-1.00 |
Specificity | 0.87 (97/111) |
95% CI | 0.80-0.95 |
Positive likelihood ratio | 7.7 |
95% CI | 4.1-14.4 |
Negative likelihood ratio | 0.03 |
95% CI | 0.01-0.09 |
SureSight Autorefractor test result, positive/negative |
|
Strabismus, amblyopia, or both | 127/44 |
Neither | 43/69 |
Sensitivity | 0.74 (127/171) |
95% CI | 0.66-0.83 |
Specificity | 0.62 (69/112) |
95% CI | 0.50-0.73 |
Positive likelihood ratio | 1.9 |
95% CI | 1.4-2.7 |
Negative likelihood ratio | 0.42 |
95% CI | 0.29-0.61 |
Abbreviation: PVS, Pediatric Vision Scanner.
Data are given as number/number unless otherwise noted.