Table 4. Screening Errors by Gold Standard Examination Diagnoses and Cycloplegic Refractiona.
Characteristic | PVS | SureSight Autorefractor |
---|---|---|
False negative | ||
Strabismus | 0 [73] | 27 (40.3) [67] |
Anisometropic amblyopia | 3 (8.1) [37] | 5b (14.7) [34] |
Bilateral amblyopia | 1 (5.0) [20] | 2 (12.5) [16] |
Strabismus + amblyopia | 1 (1.7) [58] | 10 (18.5) [54] |
Total | 5 (2.7) [188] | 44 (25.7) [171] |
False positive | ||
Hyperopia | 0 | 0 |
Myopia | 2 | 1 |
Astigmatism | 2 | 17 |
Anisometropia | 2 | 7 |
Astigmatic anisometropia | 1 | 3 |
Normal | 7 | 15 |
Total | 14 (12.6) [111] | 43 (38.4) [112] |
Abbreviation: PVS, Pediatric Vision Scanner.
Values are given as number (percentage) [number tested with each screening device in each diagnostic category].
Of the 2 children with fixation-defined amblyopia and no strabismus, the only error was a false-negative finding of anisometropic amblyopia with the SureSight Autorefractor.