Skip to main content
. 2015 Feb 12;13:13. doi: 10.1186/s12951-015-0076-6

Table 1.

Comparison of βHBA determination in serum sample by three approaches

Serum sample # Method
Chemistry analyzer [μM] Microplate reader [μM] Microfluidic biosensor [μM]
1 627 530 580
2 1224 1277 1280
3 2399 2784 2433
4 3005 3093 3051
5 5152 4541 4985
Control 0 0.03 <0