Skip to main content
. 2015 Feb 11;9:6. doi: 10.1186/s13065-015-0083-6

Table 6.

Determination of NLX using AuNPs-luminol-potassium ferricyanide SIA-injection CL detection in serum and urine

Serum samples Urine samples
Taken-log conc. Found-log conc. Recovery % Taken log conc. Found-log conc. Recovery %
9.0 8.94 99.3 9.0 9.00 100.0
8.0 7.99 99.9 8.0 7.99 99.8
7.0 6.96 99.4 7.0 6.95 99.3
6.0 5.98 99.7 6.0 5.87 97.8
5.0 4.96 99.2 5.0 4.96 99.2
4.0 3.93 98.3 4.0 3.94 98.5
3.0 2.97 99.0 3.0 2.99 99.7
Mean %±SD 99.3±0.5 99.2±0.8
n 7 7
Variance 0.25 0.64
% SE 0.19 0.30
% RSD 0.50 0.80