TABLE I.

Timetable for LC/MS/MS Gradient Elution Profiles and Switching Valves

Time (min) Shimadzu gradient for analytical column elution
Waters gradient for guard column cleanup
Switching valves
Flow rate (μl/min) Mobile phase A (%) Mobile phase B (%) Flow rate (μl/min) Mobile phase A (%) Mobile phase B (%) ID Position Flow direction
0.00 175 87 13 200 87 13 1 A To cleanup column
0.00 2 B To waste
1.00 1 B To analytical column
4.50 2 A To mass spec source
4.75 200 87 13
5.00 350 0 100
8.25 175 20 80
8.50 175 5 95 350 0 100
8.75 350 87 13
9.50 175 5 95 2 B To waste
10.0 175 87 13
14.0 1 A To cleanup column
15.0 175 87 13 350 87 13