Published in final edited form as: Prostate. 2014 May;74(7):722–731. doi: 10.1002/pros.22792

TABLE III.

Plasma Calibrator and QC Validation Results

Sample type Androgen Concentrations (ng/ml) Precision (% RSD)
Accuracy (% AR)
Min Max Mean Min Max Mean
Calibrators ASD 0.0125–3.75 0.83 6.76 3.89 93.2 106 100
T 0.00625–3.75 0.77 5.32 2.53 92.6 105 100
DHEA 0.250–7.50 1.33 7.36 3.99 98.2 104 100
DHT 0.0250–7.50 1.47 7.66 3.76 95.6 105 99.8
AND 0.500–7.50 3.32 6.26 3.93 94.2 107 99.8
QCs ASD 0.0600, 0.420, 2.94 1.77 3.07 2.42 95.6 101 99.4
T 0.0270, 0.270, 2.70 2.04 2.30 2.19 94.0 101 96.8
DHEA 0.600, 1.80, 5.40 1.96 5.79 4.28 92.9 96.5 94.6
DHT 0.0540, 0.540, 5.40 3.56 6.76 5.08 92.2 95.4 93.3
AND 0.600, 1.80, 5.40 3.66 13.4 7.75 92.6 114 106

RSD, relative standard deviation; AR, analytical recovery.