Table 5.
Average volume of droplets generated by the Bio-Rad Droplet Generator using either Supermix for Probes or EvaGreen Supermix. The uncertainty on the average copy number concentration is an expanded uncertainty with a coverage factor k of 2
| Cartridge | Mastermix | Droplet volume (nL) | Total number of droplets |
|---|---|---|---|
| DG8 (production year 2012)a | ddPCR™ Supermix for Probes | 0.833 ± 0.035 | 622 |
| DG8 (lot nos. C000021590, C000024241 and C000031616)a | QX200™ ddPCR™ EvaGreen® Supermix | 0.830 ± 0.027 | 805 |
| DG8 (production year 2013)b | ddPCR™ Supermix for Probes | 0.860 ± 0.021 | 463 |
aDroplets were analysed from three wells from each of five DG8 cartridges (15 wells in total)
bSixteen 2 μL samples were collected sequentially from droplets generated from a single well of a DG8 cartridge and the average droplet volume measured