. 2015 Jan 20;407(7):1831–1840. doi: 10.1007/s00216-015-8458-z

Table 5.

Average volume of droplets generated by the Bio-Rad Droplet Generator using either Supermix for Probes or EvaGreen Supermix. The uncertainty on the average copy number concentration is an expanded uncertainty with a coverage factor k of 2

Cartridge Mastermix Droplet volume (nL) Total number of droplets
DG8 (production year 2012)a ddPCR™ Supermix for Probes 0.833 ± 0.035 622
DG8 (lot nos. C000021590, C000024241 and C000031616)a QX200™ ddPCR™ EvaGreen® Supermix 0.830 ± 0.027 805
DG8 (production year 2013)b ddPCR™ Supermix for Probes 0.860 ± 0.021 463

aDroplets were analysed from three wells from each of five DG8 cartridges (15 wells in total)

bSixteen 2 μL samples were collected sequentially from droplets generated from a single well of a DG8 cartridge and the average droplet volume measured