Skip to main content
. Author manuscript; available in PMC: 2016 Jan 20.
Published in final edited form as: Curr Protoc Hum Genet. 2015 Jan 20;84:11.14.1–11.14.23. doi: 10.1002/0471142905.hg1114s84

Table 4.

Troubleshooting

Problem Cause Solution
Low PCR1 yield Adapter dimers in PCR1 or low starting quantities Repeat AMPure XP beads clean-up and/or amplify 5–10 additional cycles. Check the AMPure XP bead calibration.
Low PCR2 yield Adapter dimers in PCR1 and/or PCR2 or too much salt in the DSN reaction Repeat AMPure XP beads clean-up on PCR1 and/or amplify 5–10 additional cycles. Check the AMPure XP bead calibration.
Poor rRNA reduction Adapter dimers in PCR1 or mechanical failure during the DSN reaction Remove adapter dimers with additional AMPure XP bead clean ups. Mix the DSN hybridization and reaction well, keep the temperature at a constant 68°C during these steps.
No mature miRNA in the smRNA library Loss during RNeasy isolation of RNA or AMPure XP bead clean-up due to size selection or low ethanol concentration in the bead washes Use RWT buffer with the proper amount of ethanol for the RNeasy column wash. Check the AMPure XP bead calibration. Make 80% ethanol fresh for RNeasy column and bead washes.
Reduction of rRNA qRT-PCR validation inaccurate Secondary inflection in the amplification curve resulting in high CT values for the library rRNA amplification Use the ‘Absolute Quantification Analysis Using Fit Points Method’ to set the cycle count lower for qRT-PCR on the 480 LightCycler.
Poor PCA for biological replicates Poor size fractionation of the input RNA due to overloading the RNeasy Minelute column Use RNeasy mini column for the size fractionation or use less RNA. Typically the data will still be robust.