Table 4.
Troubleshooting
Problem | Cause | Solution |
---|---|---|
Low PCR1 yield | Adapter dimers in PCR1 or low starting quantities | Repeat AMPure XP beads clean-up and/or amplify 5–10 additional cycles. Check the AMPure XP bead calibration. |
Low PCR2 yield | Adapter dimers in PCR1 and/or PCR2 or too much salt in the DSN reaction | Repeat AMPure XP beads clean-up on PCR1 and/or amplify 5–10 additional cycles. Check the AMPure XP bead calibration. |
Poor rRNA reduction | Adapter dimers in PCR1 or mechanical failure during the DSN reaction | Remove adapter dimers with additional AMPure XP bead clean ups. Mix the DSN hybridization and reaction well, keep the temperature at a constant 68°C during these steps. |
No mature miRNA in the smRNA library | Loss during RNeasy isolation of RNA or AMPure XP bead clean-up due to size selection or low ethanol concentration in the bead washes | Use RWT buffer with the proper amount of ethanol for the RNeasy column wash. Check the AMPure XP bead calibration. Make 80% ethanol fresh for RNeasy column and bead washes. |
Reduction of rRNA qRT-PCR validation inaccurate | Secondary inflection in the amplification curve resulting in high CT values for the library rRNA amplification | Use the ‘Absolute Quantification Analysis Using Fit Points Method’ to set the cycle count lower for qRT-PCR on the 480 LightCycler. |
Poor PCA for biological replicates | Poor size fractionation of the input RNA due to overloading the RNeasy Minelute column | Use RNeasy mini column for the size fractionation or use less RNA. Typically the data will still be robust. |