Table 3.

Distribution of Changes in Device Variables*

Variable Patients, n Decrease, n (%)
≤20% Change, n (%) Increase, n (%)
41%–50% 31%–40% 21%–30% 21%–30% 31%–40% 41%–50%
Immediately after MRI

 P-wave amplitude 303 0 (0) 4 (1.3) 24 (7.9) 245 (80.9) 10 (3.3) 20 (6.6) 0 (0)

 Right ventricular R-wave amplitude 461 0 (0) 5 (1.1) 19 (4.1) 419 (90.9) 7 (1.5) 11 (2.4) 0 (0)

 Left ventricular R-wave amplitude 20 0 (0) 1 (5) 3 (15) 15 (75) 1 (5) 0 (0) 0 (0)

 Atrial lead impedance 364 0 (0) 0 (0) 0 (0) 363 (99.7) 1 (0.3) 0 (0) 0 (0)

 Right ventricular lead impedance 518 0 (0) 0 (0) 1 (0.2) 515 (99.4) 2 (0.4) 0 (0) 0 (0)

 Left ventricular lead impedance 51 0 (0) 0 (0) 1 (2) 50 (98) 0 (0) 0 (0) 0 (0)

 Atrial capture threshold 332 15 (4.5) 9 (2.7) 13 (3.9) 265 (79.8) 7 (2.1) 13 (3.9) 10 (3.0)

 Right ventricular capture threshold 531 11(2.1) 15 (2.8) 13 (2.5) 455 (85.7) 9 (1.7) 15 (2.8) 13 (2.5)

 Left ventricular capture threshold 55 0 (0) 2 (4) 3 (5) 44 (80) 4 (7) 2 (4) 0 (0)

 Battery voltage 472 0 (0) 0 (0) 0 (0) 472 (100.0) 0 (0) 0 (0) 0 (0)
Long-term follow-up after MRI

 P-wave amplitude 124 0 (0) 6 (5) 10 (8) 94 (76) 6 (5) 8 (6) 0 (0)

 Right ventricular R-wave amplitude 198 0 (0) 15 (8) 11 (6) 155 (78) 8 (4) 9 (5) 0 (0)

 Left ventricular R-wave amplitude 9 0 (0) 1 (11) 1 (11) 7 (78) 0 (0) 0 (0) 0 (0)

 Atrial lead impedance 167 0 (0) 0 (0) 2 (1) 162 (97) 3 (2) 0 (0) 0 (0)

 Right ventricular lead impedance 234 0 (0) 0 (0) 3 (1) 223 (95) 8 (3) 0 (0) 0 (0)

 Left ventricular lead impedance 23 0 (0) 0 (0) 0 (0) 23 (100) 0 (0) 0 (0) 0 (0)

 Atrial capture threshold 105 1 (1) 2 (2) 6 (6) 86 (82) 3 (3) 3 (3) 4 (4)

 Right ventricular capture threshold 190 7 (4) 10 (5) 8 (4) 122 (64) 7 (4) 16 (8) 20 (11)

 Left ventricular capture threshold 23 2 (9) 2 (9) 1 (4) 10 (43) 1 (4) 4 (17) 3 (13)
 Battery voltage 224 0 (0) 0 (0) 3 (1) 221 (99) 0 (0) 0 (0) 0 (0)

MRI = magnetic resonance imaging.

*

Lead impedance variations exceeding 30% (20, 21), capture threshold variations exceeding 50% (22), and sensing variations exceeding 40% (23, 24) indicate clinically significant changes in lead performance.

The number of comparisons for each device variable is unique, primarily owing to variability in the number of leads, the absence of intrinsic P/R waves, the presence of atrial arrhythmia, and pulse widths during measurement of capture threshold at follow-up interrogation. The proportion of missing data for each variable is discussed in the Discussion section.

Based on long-term follow-up results available for 266 (61%) patients undergoing 278 MRI examinations.