Table 2.

Comparison of the OI-RD/microarray-based assay platform with Biacore 3000 in characterization of specific affinity constants and rate constants of monoclonal rabbit antibodies with 6 immobilized peptide targets.

Biacore 3000 OI-RD scanning sensor

Original proteins from which 6 peptides are derived kon 105s−1M−1 koff 10−5s−1 Kd pM kon 105s−1M−1 koff 10−5s−1 Kd pM
CD34 3.5 1.4 41 0.37 0.43 115
Cytokeratin 15 7.3 1.8 247 0.16 0.89 564
EGFR Phospho 0.69 0.058 3.5 0.42 0.08 19
Glucagon 3.7 2.6 70 0.35 1.2 331
Glut-1 4.6 1.1 24 0.36 0.028 7.7
Vimentin 2.3 7.1 306 0.25 4.3 1700