Quantitative EEG Analysis for Automated Detection of Nonconvulsive Seizures in Intensive Care Units

J Chris Sackellares a,*, Deng-Shan Shiau a, Jonathon J Halford b, Suzette M LaRoche c, Kevin M Kelly d,e
PMCID: PMC4342615  NIHMSID: NIHMS322211  PMID: 22078521

Abstract

Due to increased awareness of the high prevalence of nonconvulsive seizures (NCSs) in critically ill patients, continuous EEG monitoring (cEEG) in ICUs is rapidly increasing in use. However, cEEG monitoring is labor intensive; manual review and interpretation of the EEG are impractical in most ICUs. Effective methods to assist in rapid and accurate detection of NCSs would greatly reduce the cost of cEEG and enhance the quality of patient care. In this study, we report a preliminary investigation of a novel ICU EEG analysis and seizure detection algorithm. Twenty-four prolonged cEEG recordings were included in this study. Seizure detection sensitivity and specificity were assessed for the new algorithm and for the two commercial seizure detection software systems.

The new algorithm performed a mean sensitivity of 90.4% and a mean false detection rate of 0.066/h. The two commercial detection products performed with low sensitivities (12.9% and 10.1%) and false detection rates of 1.036/h and 0.013/h, respectively.

These findings suggest that the novel algorithm has potential to be the basis of clinically useful software that can assist ICU staff in timely identification of NCSs. This study also suggests that currently available seizure detection software does not have sufficient performance for the detection of NCSs in critically ill patients.

Keywords: ICU, Nonconvulsive seizures, cEEG monitoring, Seizure detection, Quantitative EEG trending

Introduction

Seizures occur as a consequence of a variety of acute insults to the brain [1-3]. These insults include CNS infections, intracranial hemorrhage, ischemic stroke, traumatic brain injury, and brain tumor. Seizures also occur as a consequence of toxic and metabolic encephalopathies. As a result, seizures are frequently encountered in critical care settings. Some seizures manifest as convulsions and can be easily recognized when a trained observer is nearby. However, 92% of seizures that occur in critically ill patients are nonconvulsive [4-9]. Because nonconvulsive seizures cannot be diagnosed with certainty on the basis of clinical manifestations, continuous EEG (cEEG) monitoring is required for reliable detection [4-9]. The high incidence of discrete nonconvulsive seizures, as well as sustained nonconvulsive status epilepticus, particularly in patients with intracranial hemorrhage and recent convulsive status epilepticus, has been documented by several investigators [3-4, 7, 10-14]. Without timely treatment, nonconvulsive seizures may impair consciousness and may lead to secondary brain injury and worse clinical outcome [3-5, 8-10, 12, 14-18]. As a result, there is growing recognition of the need to monitor the EEG in patients treated in critical care units.

The utility of cEEG monitoring in ICUs has been well established [8, 19-26]. However, most ICUs are not staffed by professionals with expertise in neurology and clinical neurophysiology. As a result, in contrast to other vital organ systems, routine continuous monitoring of brain activity is not performed in most ICUs. Even in hospitals staffed with experts in neurology and electroencephalography, cEEG monitoring is labor-intensive and continuous review and interpretation of raw EEG data are impractical in most clinical settings. Tools that aid rapid and accurate detection of NCSs would greatly improve the efficiency of cEEG monitoring and enhance the quality of patient care. An automated, reliable, and user-friendly system for detecting critical changes in the EEG, such as NCSs, could help make EEG monitoring more practical in ICUs. In addition, a mechanism for simultaneously transmitting the EEG to remote sites for interpretation is highly desirable.

Automated seizure detection software is often used to assist in EEG review. However, in the ICU setting, seizures are caused by a wide range of neurological and systemic disorders. As a result, electrographic seizure patterns can differ substantially from seizures recorded in patients with chronic epilepsy in outpatient and EMU settings. In many instances in the ICU, the electrographic patterns of NCSs differ significantly from patterns that occur in epileptic patients. Some NCS patterns strongly resemble other organized rhythmic patterns in the EEG, such as the triphasic waves seen in metabolic encephalopathies [27]. As a result, use of currently available seizure detection software in ICUs results in either low sensitivity or high false detections, or both, and therefore does not offer a detection performance that is clinically useful. There are few published articles reporting on the utility of automated algorithms for detecting seizures in acute care settings. A preliminary study by Zhang and colleagues [28] suggests that it is possible to develop automated algorithms that can distinguish between nonconvulsive seizure patterns that occur in acute care settings and other organized rhythmic patterns characteristic of toxic or metabolic encephalopathies. A workshop held by the FDA (Title: Seizure Detection, Cognitive Function, and TBI/Concussion Devices: Issues in Their Evaluation) recently discussed the need of validating seizure detection algorithms intended to be used in ICU patients (http://www.regonline.com/builder/site/Default.aspx?EventID=960009).

Because of the inadequacy of seizure detection software in the ICU, “EEG trends” based on quantitative EEG (qEEG) analysis are often used in the ICU setting to facilitate detection of NCS patterns in long-term EEG recordings. To generate EEG trends, EEG waveforms are transformed into multi-channel graphic representation of quantitative EEG measures, such as signal amplitude and/or frequency. The trends are commonly presented in windows of 30 to 120 minutes, making it easier to detect clinically significant transient events, such as seizure discharges, by visual inspection of the trends. Potentially, EEG trending can facilitate the EEG review process. However, EEG trending software is limited because it has difficulty representing brief, focal or slowly evolving, low frequency seizures (Fig. 1) in a form that is easily recognizable. Trends can be a useful aid to experienced electroencephalographers and EEG technicians. However, to our knowledge, there is no published study that evaluates the clinical utility of trending in acute care settings.

Figure 1.

Figure 1

A 4-channel Envelope trend over a 2-hour period containing a seizure pattern of about 3 minutes (upper trace) shows little change during a low-frequency NCS pattern that is present in the original EEG signal (lower trace shows 11 seconds of the ictal discharge).

While existing EEG-trending software appears to be much more useful than current seizure detection software in the Neuro-Critical Care setting, there are significant technical limitations to existing methods. One of the most useful EEG trends is the “Envelope” trend. The Envelope trend selects waveforms within a specific frequency range and plots the median waveform amplitude over a specified period of time (usually 10 to 20 seconds). However, as shown in Figure 1, this trend often does not facilitate detection of some NCS patterns.

Development of an effective automated seizure detection algorithm for use in acute care settings remains a challenge for two primary reasons. First, the algorithm must detect a wide variety of ictal EEG patterns, many of which are not seen in patients with chronic epilepsy. Second, the algorithm must reject organized rhythmic patterns produced by sleep, anesthesia, and encephalopathies as well as repetitive organized or rhythmic patterns produced by artifacts commonly present in long term recordings in acute care settings.

In this study, we report a preliminary evaluation of a novel ICU automated seizure detection algorithm (ICU-ASDA). ICU-ASDA utilizes novel measures of signal characteristics that can characterize essentially all ictal EEG patterns. Based on these measures, ICU-ASDA incorporates specially designed detection and rejection criteria. The algorithm is illustrated in Figure 2. The performance of ICU-ASDA was compared to that of two commercially available detection programs.

Figure 2.

Figure 2

Flow chart of the proposed ICU seizure detection algorithm

Methods

ICU recordings

The ICU-ASDA was developed and tested in a sample of 24 long-term continuous ICU EEG recordings obtained previously for clinical diagnostic purposes at the Medical University of South Carolina and Emory University Hospital. The total duration of these 24 recordings is approximately 1,500 hours, with a total of 453 NCSs in 11 recordings. Datasets were not preselected before the analysis and therefore included ICU patients with a variety of etiologies and ictal EEG patterns. Seizure events were determined and documented at the clinical sites, with an additional independent review by an additional electroencephalographer (JCS).

The Algorithm

The algorithm introduced here uses signal amplitude variation (AV) for detecting significant changes in amplitude and a regularity statistic (PMRS), based on the repeatability of the signal patterns [29-31] for detecting rhythmicity changes. The algorithm also calculates maximum signal frequency (Fmax) and amplitudes in a high frequency range (>25 Hz; Ahf (amplitude in a high frequency range)), for rejecting unwanted detections (e.g., sleep transients, muscle and chewing artifacts). Although different montages can be applied, analysis was performed using a 16-channel referential montage that included electrodes at Fp1, Fp2, F3, F4, C3, C4, P3, P4, O1, O2, F7, F8, T3, T4, T 5, and T6, referenced to an electrode located between Cz and Pz, as suggested by the American Clinical Neurophysiology Society [32]. The algorithm detects the presence of seizure discharges, and recognizes three onset patterns (i.e., left-unilateral, right-unilateral, or bilateral).

The steps of the algorithm are depicted in Figure 2. Signals are filtered with a 5th order Butterworth filter to reduce unwanted detections due to background activity in the very low and high frequency ranges. After filtering the signals, the algorithm calculates several qEEG measures (PMRS, AV, Fmax, and Ahf) for each of the EEG channels (steps 3a and 3b). Based on these qEEG measurements, step 4 adopts three artifact rejection criteria (ARC) to eliminate false detections. If an EEG window is rejected, the algorithm starts over from step 1 and processes the next EEG epoch. Otherwise, the algorithm goes on to step 6 to determine if the EEG epoch satisfies one of the three detection criteria: (1) left-unilateral ictal pattern, (2) right-unilateral ictal pattern, or (3) bilateral ictal pattern. Figure 3 gives an example of a NCS with a left-unilateral ictal pattern (low-amplitude discharge over the left occipital/parietal region) that was detected by the algorithm.

Figure 3.

Figure 3

Demonstration of a low-amplitude ICU nonconvulsive seizure (left occipital/parietal) detected by the test ICU seizure detection algorithm

Statistical Analysis

Each of the test EEG recordings was processed by the test algorithm (with fixed thresholds of detection/rejection criteria) and two currently available seizure detection software systems on the market: Persyst’s Reveal and Optima’s IdentEvent. Detection performance (sensitivity and the false detection rate per hr) of each detection algorithm was calculated. In order to prevent over-emphasizing overall performance on recordings with longer durations or larger number of seizures, sensitivity and the false detection rate were calculated for each individual recording and the mean values were used as the final overall performance.

Results

Table I provides details of data characteristics as well as the detection performance by the three algorithms. Overall, the ICU-ASDA performed at a mean sensitivity of 90.4% over the 11 recordings that included identified seizure, and a mean false detection rate of 0.066 per hour (i.e., 1.6 per 24 hours) over all 24 studied recordings. This encouraging training performance gives us high confidence to proceed to a formal performance validation study. It is worth noting that, over the same datasets, the two currently commercially available seizure detection software offer clearly sub-par detection performance for detecting NCSs in ICU patients. We believe that this is the reason why the FDA has recently highlighted the importance of independent (from EMU patients) performance validation of seizure detection devices for use in ICU settings, which has not been reported in the literature.

Table I.

Data characteristics and detection performance statistics

Detection Performance Statistics
Recording Duration (h) # of Seizures ICU-ASDA Reveal IdentEvent
Sensitivity FDR / hr Sensitivity FDR / hr Sensitivity FDR / hr
01 12.29 10 1.000 0.000 0.300 1.058 0.000 0.000
02 8.19 25 0.800 0.000 0.004 0.611 0.000 0.000
03 6.82 28 0.929 0.147 0.107 3.079 0.000 0.000
04 3.40 26 0.923 0.000 0.000 0.294 0.000 0.000
05 23.39 0 0.171 2.095 0.000
06 42.06 0 0.048 0.143 0.000
07 55.53 0 0.072 1.765 0.018
08 44.58 0 0.067 0.606 0.000
09 28.73 0 0.070 0.070 0.000
10 311.05 0 0.100 0.884 0.003
11 63.95 0 0.000 3.018 0.016
12 44.37 0 0.000 0.135 0.000
13 19.84 0 0.000 0.000 0.050
14 278.17 50 0.940 0.151 0.200 0.417 0.360 0.007
15 25.67 0 0.078 0.156 0.000
16 21.93 0 0.091 0.365 0.000
17 27.06 5 0.800 0.037 0.000 0.222 0.000 0.000
18 27.46 0 0.146 6.664 0.000
19 138.55 18 1.000 0.007 0.111 0.144 0.222 0.000
20 47.42 16 0.750 0.148 0.000 0.042 0.063 0.021
21 15.27 217 0.876 0.000 0.055 0.000 0.028 0.000
22 49.11 5 1.000 0.020 0.400 0.224 0.000 0.000
23 162.14 0 0.148 0.703 0.074
24 43.55 53 0.925 0.092 0.208 2.158 0.434 0.115
Total/Overall 1,500.53 453 90.4% 0.066 / h = 1.6 /day 12.9% 1.036 /h = 24.9 / day 10.1% 0.013 / h = 0.3 / day

Discussion

There is growing recognition that EEG brain monitoring in acute care settings detects seizures and other acute causes of disturbed brain function that would not be detected otherwise. As a result, EEG monitoring in ICU settings and in Emergency Departments is increasing. This has led to the recognition that seizures are much more common in acutely ill patients than had been suspected, based on clinical examination and routine 20 to 30 minute EEG recordings. Early detection of seizures may lead to interventions that can reduce mortality and morbidity, although further study is needed to establish this. Increasing usage of continuous EEG monitoring can place excessive demands on the time of electroencephalographers - even when assisted by highly trained technicians - to interpret the large quantities of EEG data produced. Thus, there is a need to develop accurate diagnostic tools that will improve accuracy of detection and increase efficiency of record review.

Currently available commercial seizure detection software has been developed for and tested on recordings obtained from patients with uncontrolled epileptic seizures generated in EMUs. Because of major differences in EEG background patterns, artifact sources and ictal EEG patterns, these detectors do not perform well in acute care settings. This study demonstrates the poor performance of currently available commercial seizure detectors in recordings obtained from the ICU. This observation is not surprising, given the fact that the algorithms were not developed for this application. There is a need to develop seizure detection methods specifically designed for use in acute care settings. We report a preliminary study to evaluate a novel seizure detector algorithm developed for such a setting, the ICU-ASDA. These preliminary findings suggest that it is possible to develop algorithms with sufficient sensitivity and specificity for use in acute clinical settings

  • Use of EEG monitoring is increasing in ICUs to identify nonconvulsive seizures.

  • No clinically validated seizure detection algorithms exist for ICU patients.

  • A new seizure detection algorithm was tested in 24 prolonged EEG recordings.

  • The new algorithm performed 90.4% sensitivity and 0.066/h false detection rate.

  • This suggests that the algorithm can assist ICU staff in identifying seizures.

Footnotes

Publisher's Disclaimer: This is a PDF file of an unedited manuscript that has been accepted for publication. As a service to our customers we are providing this early version of the manuscript. The manuscript will undergo copyediting, typesetting, and review of the resulting proof before it is published in its final citable form. Please note that during the production process errors may be discovered which could affect the content, and all legal disclaimers that apply to the journal pertain.

Contributor Information

Dr. Deng-Shan Shiau, Email: dshiau@optimaneuro.com.

Dr. Jonathon J. Halford, Email: halfordj@musc.edu.

Dr. Suzette M. LaRoche, Email: slaroch@emory.edu.

Dr. Kevin M. Kelly, Email: kelly@wpahs.org.

References

  • 1.Young GB, Jordan KG, Doig GS. An assessment of nonconvulsive seizures in the intensive care unit using continuous EEG monitoring: An investigation of variables associated with mortality. Neurology. 1996;47(1):83–89. doi: 10.1212/wnl.47.1.83. [DOI] [PubMed] [Google Scholar]
  • 2.Waterhouse EJ, Vaughan JK, Barnes TY, Boggs JG, Towne AR, Kopec-Garnett L, DeLorenzo RJ. Synergistic effect of status epilepticus and ischemic brain injury on mortality. Epilepsy Res. 1998;29:175–183. doi: 10.1016/s0920-1211(97)00071-5. [DOI] [PubMed] [Google Scholar]
  • 3.Jirsch J, Hirsch LJ. Nonconvulsive seizures: Developing a rational approach to the diagnosis and management in the critically ill population. Clin Neurophysiol. 2007;118:1660–1670. doi: 10.1016/j.clinph.2006.11.312. [DOI] [PubMed] [Google Scholar]
  • 4.Jordan KG. Continuous EEG and evoked potential monitoring in the neuroscience intensive care unit. J Clin Neurophysiol. 1993;10:445–475. doi: 10.1097/00004691-199310000-00006. [DOI] [PubMed] [Google Scholar]
  • 5.Vespa PM, Nuwer MR, Nenov V, Ronne-Engstrom E, Hovda DA, Bergsneider M, Kelly DF, Martin NA, Becker DP. Increased incidence and impact of nonconvulsive and convulsive seizures after traumatic brain injury as detected by continuous EEG in the intensive care unit. J Neurosurg. 1999;91(5):750–760. doi: 10.3171/jns.1999.91.5.0750. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Claassen J, Bernardini GL, Kreiter KT, Bates JE, Du YE, Copeland D, Connolly ES, Mayer SA. Effect of cisternal and ventricular blood on risk of delayed cerebral ischemia after subarachnoid hemorrhage: the Fisher Scale revisited. Stroke. 2001;32:2012–2020. doi: 10.1161/hs0901.095677. [DOI] [PubMed] [Google Scholar]
  • 7.Vespa PM, McArthur D, O’Phelan K, Glenn T, Etchepare M, Kelly D, Bergsneider M, Martin NA, Hovda DA. Persistently low extracellular glucose correlates with poor outcome 6 months after human traumatic brain injury despite a lack of increased lactate: a microdialysis study. J Cereb Blood Flow Metab. 2003;23:865–877. doi: 10.1097/01.WCB.0000076701.45782.EF. [DOI] [PubMed] [Google Scholar]
  • 8.Claassen J, Mayer SA, Kowalski RG, Emerson RG, Hirsch LJ. Detection of electrographic seizures with continuous EEG monitoring in critically ill patients. Neurology. 2004;62:1743–1748. doi: 10.1212/01.wnl.0000125184.88797.62. [DOI] [PubMed] [Google Scholar]
  • 9.Pandian JD, Cascino GD, So EL, Manno E, Fulgham JR. Digital video electroencephalographic monitoring in the neurological-neurosurgical intensive care unit: clinical features and outcome. Arch Neurol. 2004;61:1090–1094. doi: 10.1001/archneur.61.7.1090. [DOI] [PubMed] [Google Scholar]
  • 10.Hosain SA, Solomon GE, Kobylarz EJ. Electroencephalographic patterns in unresponsive pediatric patients. Pediatr Neurol. 2005;32:162–165. doi: 10.1016/j.pediatrneurol.2004.09.008. [DOI] [PubMed] [Google Scholar]
  • 11.Claassen J, Jette N, Chum F, Green R, Schmidt M, Choi H, Jirsch J, Frontera JA, Sander Connolly E, Emerson RG, Mayer SA, Hirsch LJ. Electrographic seizures and periodic discharges after intracerebral hemorrhage. Neurology. 2007;69:1356–1365. doi: 10.1212/01.wnl.0000281664.02615.6c. [DOI] [PubMed] [Google Scholar]
  • 12.DeLorenzo RJ, Waterhouse EJ, Towne AR, Boggs JG, Ko D, DeLorenzo GA, Brown A, Garnett L. Persistent nonconvulsive status epilepticus after the control of convulsive status epilepticus. Epilepsia. 1998;39:833–840. doi: 10.1111/j.1528-1157.1998.tb01177.x. [DOI] [PubMed] [Google Scholar]
  • 13.Treiman DM, Meyers PD, Walton NY, Collins JF, Colling C, Rowan AJ, Handforth A, Faught E, Calabrese VP, Uthman BM, Ramsay RE, Mamdani MB, Yagnik P, Jones JC, Barry E, Boggs JG, Kanner AM. The Veterans Affairs Status Epilepticus Cooperative Study Group. A Comparison of Four Treatments for Generalized Convulsive Status Epilepticus. New Engl J Med. 1998;339:792–798. doi: 10.1056/NEJM199809173391202. [DOI] [PubMed] [Google Scholar]
  • 14.Towne AR, Waterhouse EJ, Boggs JG, Garnett LK, Brown AJ, Smith JR, Jr, DeLorenzo RJ. Prevalence of nonconvulsive status epilepticus in comatose patients. Neurology. 2000;54:340–345. doi: 10.1212/wnl.54.2.340. [DOI] [PubMed] [Google Scholar]
  • 15.Friedman D, Claassen J, Hirsch LJ. Continuous electroencephalogram monitoring in the intensive care unit. Anesth Analg. 2009;109:506–523. doi: 10.1213/ane.0b013e3181a9d8b5. [DOI] [PubMed] [Google Scholar]
  • 16.Jette N, Claassen J, Emerson RG, Hirsch LJ. Frequency and predictors of nonconvulsive seizures during continuous electroencephalographic monitoring in critically ill children. Arch Neurol. 2006;63:1750–1755. doi: 10.1001/archneur.63.12.1750. [DOI] [PubMed] [Google Scholar]
  • 17.Hyllienmark L, Amark P. Continuous EEG monitoring in a paediatric intensive care unit. Eur J Paediatr Neurol. 2007;11:70–75. doi: 10.1016/j.ejpn.2006.11.005. [DOI] [PubMed] [Google Scholar]
  • 18.Oddo M, Carrera E, Claassen J, Mayer SA, Hirsch LJ. Continuous electroencephalography in the medical intensive care unit. Critical Care Medicine. 2009;37:2051–2056. doi: 10.1097/CCM.0b013e3181a00604. [DOI] [PubMed] [Google Scholar]
  • 19.Jordan KG. Nonconvulsive status epilepticus in acute brain injury. J Clin Neurophysiol. 1999;16(4):332–340. doi: 10.1097/00004691-199907000-00005. [DOI] [PubMed] [Google Scholar]
  • 20.Hirsch LJ. Continuous EEG monitoring in the intensive care unit: An overview. J Clin Neurophysiol. 2004;21(5):332–340. [PubMed] [Google Scholar]
  • 21.Hirsch LJ, Kull LL. Continuous EEG Monitoring in Intensive Care Unit. Am J Electroneurodiagnostic Technol. 2004;44:137–158. [PubMed] [Google Scholar]
  • 22.Hirsch LJ. Brain monitoring: The next frontier of ICU monitoring. J Clin Neurophysiol. 2004a;21(5):305–306. [PubMed] [Google Scholar]
  • 23.Hirsch LJ. Continuous EEG monitoring in the intensive care unit: An overview. J Clin Neurophysiol. 2004b;21(5):332–340. [PubMed] [Google Scholar]
  • 24.Kaplan PW. The EEG of status epilepticus. J Clin Neurophysiol. 2006;23(3):221–229. doi: 10.1097/01.wnp.0000220837.99490.66. [DOI] [PubMed] [Google Scholar]
  • 25.Kull LL, Emerson RG. Continuous EEG monitoring in the intensive care unit: Technical and staffing considerations. J Clin Neurophysiol. 2005;22(2):107–118. doi: 10.1097/01.wnp.0000158361.24544.2d. [DOI] [PubMed] [Google Scholar]
  • 26.Scheuer ML, Wilson SB. Data analysis for continuous EEG monitoring in the ICU: Seeing the forest and the trees. J Clin Neurophysiol. 2004;21(5):353–378. [PubMed] [Google Scholar]
  • 27.Bearden S, Eisenschenk S, Uthman B. Diagnosis of nonconvulsive status epilepticus (NCSE) in adults with altered mental status: clinico-electroencephalographic considerations. Am J Electroneurodiagnostic Technol. 2008;48(1):11–37. [PubMed] [Google Scholar]
  • 28.Zhang J, Xanthopoulos P, Liu CC, Bearden S, Uthman BM, Pardalos PM. Real-time differentiation of nonconvulsive status epilepticus from other encephalopathies using quantitative EEG analysis: a pilot study. Epilepsia. 2009;51(2):243–250. doi: 10.1111/j.1528-1167.2009.02286.x. [DOI] [PubMed] [Google Scholar]
  • 29.Shiau DS, Halford J, Kelly KM, Kern RT, Inman M, Chien JH, Pardalos PM, Yang MCK, Sackellares JC. Signal Regularity-Based Automated Seizure Detection System for Scalp EEG Monitoring. Cybernet Syst Analysis. 2010;46(6):922–935. doi: 10.1007/s10559-010-9273-3. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 30.Kelly KM, Shiau DS, Kern R, Chien JH, Yandora K, Valeriano J, Halford J, Sackellares JC. Assessment of a Scalp EEG-based automated seizure detection system. Clin Neurophysiol. 2010;121:1832–1843. doi: 10.1016/j.clinph.2010.04.016. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 31.Halford JJ, Shiau DS, Kern RT, Stroman C, Kelly KM, Sackellares JC. Seizure Detection Software Used to Complement the Visual Screening Process for Long-Term EEG Monitoring. Am J Electroneurodiagnostic Technol. 2010;50(2):133–147. [PMC free article] [PubMed] [Google Scholar]
  • 32.American Clinical Neurophysiology Society. Guideline 14: Guidelines for recording clinical EEG on digital media. J Clin Neurophysiol. 1994;11(1):114–115. [PubMed] [Google Scholar]

RESOURCES