. 2015 Feb 27;16(2):229–238. doi: 10.3348/kjr.2015.16.2.229

Table 1. Performance of Three Readers for Characterization of Breast Lesions at DBT and US.

Az Value Sensitivity* (%) Specificity* (%)
DBT US 95% CI P DBT US P DBT US P
Reader 1 0.896 0.935 -0.014, 0.094 0.150 98.7 (74/75) 100 (75/75) 1.000 36.4 (16/44) 43.2 (19/44) 0.581
Reader 2 0.917 0.905 -0.038, 0.062 0.648 100 (75/75) 98.7 (74/75) 1.000 25.0 (11/44) 31.8 (14/44) 0.549
Reader 3 0.932 0.911 -0.036, 0.077 0.480 93.3 (70/75) 97.3 (73/75) 0.453 72.7 (32/44) 43.2 (19/44) 0.004
Overall 0.899 0.914 -0.019, 0.049 0.394 97.3 98.7 0.508 44.7 39.4 0.360

Note.- *Data in parentheses were used to calculated percentages. Optimal cut-off, between Breast Imaging Reporting and Data System category 3 and 4A was used to yield maximal sum of sensitivity and specificity in both modality, CI = confidence interval for difference between DBT and US. DBT = digital breast tomosynthesis, US = ultrasound