. 2015 Feb 23;9:379–392. doi: 10.2147/OPTH.S75812

Table 3.

Accuracy comparison between ASA and Triple-A profiles

Parameter Time points ASA (n=107 eyes) Triple-A (n=79 eyes) P-value
% with SE within ±0.5 D (%) 1 mo 93 90 0.4215a
3 mo 94 92 0.7641a
6 mo 89 90 1.0000a
1 yr 85 90 0.3820a
% with SE within ±1.0 D (%) 1 mo 97 100 0.2630a
3 mo 99 100 1.0000a
6 mo 99 100 1.0000a
1 yr 97 100 0.2630a
P (mean) P (variance)
Postoperative SE – target (mean ± SD) 1 mo 0.11±0.40 D 0.10±0.32 D 0.731b 0.784c
3 mo 0.04±0.36 D 0.03±0.33 D 0.760b 0.427c
6 mo −0.05±0.41 D −0.01±0.34 D 0.591b 0.062c
1 yr −0.09±0.46 D −0.03±0.32 D 0.518d 0.007c

Notes:

a

Statistical analysis between the Triple-A and ASA groups was performed using Fisher’s exact test

b

two-sided Wilcoxon two-sample test

c

two-sided Siegel–Tukey test

d

two-sided median two-sample test.

Abbreviations: ASA, aberration smart ablation; SE, spherical equivalent; SD, standard deviation.