Table 1.

Technical parameters for the 3 tesla (T) and 1.5T imaging protocols utilized in this study. Note that the 3T protocol achieves 41% greater in plane spatial resolution than the 1.5T protocol in a similar overall scan time.

Parameter 3 tesla 1.5 tesla
Plane Axial Axial
Breast coil type 16 channel 8 channel
Mode 3D 3D
Sequence type Fast gradient echo Fast gradient echo
Fat Suppression SPAIR SPIR
Parallel Imaging Factor 2.7 R/L, 2.0 S/I 1.5 R/L
TR (msec) 5.9 5.6
TE (msec) 3 3
Flip angle (degrees) 10 10
Field of View (cm2) 22 A/P × 33 R/L 36 A/P × 36 R/L
Matrix 440 × 660 420 × 420
In-plane voxel size (mm) 0.5 0.85
Slice thickness (mm) 1.3 → 0.65 recon 1.6
Scan time (min) 2:51 2:53

Abbreviations: SPAIR = Spectral Attenuated Inversion Recovery, SPIR = Spectral Presaturation with Inversion Recovery, R = right, L = left, A = anterior, P= posterior, recon = reconstructed