Table 1. Particles lost and released after simulated use in in vitro vessel model.

Results Statistical Analysis
Particle Size [μm] Cetpyrsal/PTX-DCB HA/PTX-DCB PVP/PTX-DCB Cetpyrsal/PTX- vs. HA/PTX-DCB Cetpyrsal/PTX-vs. PVP/PTX-DCB HA/PTX- vs. PVP/PTX-DCB
Number of particles lost during track [#/mm²] ≥10 100 ± 106 418 ± 198 790 ± 191 ** ** *
≥25 16 ± 12 25 ± 9 53 ± 11 n.s. ** **
Number of particles transferred after dilatation [#/mm²] ≥10 70 ± 37 37 ± 7 93 ± 53 n.s. n.s. n.s.
≥25 18 ± 10 3 ± 1 12 ± 7 ** n.s. **

Number of particles (≥10 μm, ≥25 μm) per mm² balloon surface released from differently coated DCB after simulated use in an in vitro vessel model according to ASTM F2394-07, X.2.4. Values show mean ± SD of 5 experiments

n.s.: not significant

* significant: p<0.05

** very significant: p<0.01