Comparison of central corneal thickness (CCT) and the thinnest corneal thickness (TCT) readings by the Scheimpflug-Placido topographic modeling system (TMS-5) and ultrasound pachymetry.

Parameters Mean difference ± SD P value Lower 95% LoA (95% CI) Upper 95% LoA (95% CI)
CCT (μm) −15.97 ± 5.33 <0.001 −26.42 (−28.53 to −24.33) −5.52 (−7.63 to −3.42)
TCT (μm) −20.32 ± 5.28 <0.001 −30.67 (−32.75 to −28.58) −9.97 (−12.06 to −7.89)

SD: standard deviation; LoA: limits of agreement; CI: confidence interval.