Table 2.

Head-to-head comparison of S-ICD versus transvenous (TV) ICD.

  S-ICD TV-ICD
Leads within heart and vessels NO YES
Implant complications Low/Negligible Significant
X-Ray exposure NO YES
Infections Lower risk More simple to manage Higher risk Difficult to manage
Shock induced myocardial damage Negligible Significant
Patients suitable for implant About 90% Virtually all
Inappropriate shocks  4-25% (TWOS or external noise) 20-30% (supraventricular tachycardia)
Time to shock delivery 14-20” 7-9”
Pacing capability NO YES
Antitachycardia pacing NO YES
Pulse generator size 69 cc/145 grams 30 cc/70 grams
Battery life span Up to 5 years Up to 10 years
Costs High Medium/Low
Home Monitoring NO YES
Atrial arrhythmias monitoring NO YES