Table 2.

Sizing of the SAPIEN 3 Valve in Regards to the 3D Annular Area by CT as Recommended in the Edwards SAPIEN 3 Information for Use

3D Annular Area (mm2) % Annular Area Over or Under (−) Nominal by 3D CT
23 mm 26 mm 29 mm
314 29.3
320 26.9
330 23.0
338 20.1
346 17.3
350 16.0
360 12.8
370 9.7
380 6.8
390 4.1
400 1.5 29.8
410 −1.0 26.6
415 −2.2 25.1
420 −3.3 23.6
430 −5.6 20.7
440 −7.7 18.0
450 −9.8 15.3
452 14.8
460 12.8
470 10.4
480 8.1
490 5.9
500 3.8 29.8
510 1.8 27.3
520 −0.2 24.8
530 −2.1 22.5
540 −3.9 20.2
546 −4.9 18.9
550 −5.6 18.0
560 −7.3 15.9
570 −8.9 13.9
580 11.9
590 10.0
600 8.2
610 6.4
615 5.5
620 4.7
630 3.0
640 1.4
650 −0.2
660 −1.7
670 −3.1
680 −4.6
683 −5.0
690 −5.9
700 −7.3
710 −8.6
720 −9.9