Table III.

Gamma passing rates of the 3D reconstructions of the prostate plan

Reconstruction Γ

Gamma
requirement
Uniform background Back-projection Profile-based
without correction
Profile-based
with correction
3%/3 mm 96.9% 97.5% 99.4% 99.8%
2%/2 mm 93.8% 94.9% 98.5% 99.1%
1%/1 mm 78.9% 81.1% 92.7% 94.8%