Table 2.

Summary of acoustic parameters. F-number is defined as the ratio of focal depth to active aperture width and describes the imaging system focal properties. These parameters are necessary for generation of acoustic force and accurate displacement tracking for shear wave estimation.

Parameter Value
Probe P128
Push Frequency (MHz) 6.15
Track Frequency (MHz) 7.27
Push Cycles 400
Pulse Duration 65μs
Track Pulse Repetition 10330Hz
Frequency
Push F-number 1.0
Track F-number 1.5
Push focal depth 0.6cm
MI 1.9