TABLE 3.

Device performance using serum for all sites, stratified by days p.s.o.a

Day p.s.o. Test article No. of specimens that were:
Sensitivity (95% CI) (%) Specificity (95% CI) (%)
TP FN FP TN
0–3 Panbio RDT 120 130 213 371 48.0 (41.7–54.4) 63.5 (59.5–67.4)
4–8 145 17 56 90 89.5 (83.7–93.8) 61.6 (53.2–69.6)
9–14 64 1 9 17 98.5 (91.7–100.0) 65.4 (44.3–82.8)
15+ 248 5 242 339 98.0 (95.5–99.4) 58.3 (54.2–62.4)
4–14 209 18 65 107 92.1 (87.8–95.2) 62.2 (54.5–69.5)
0–3 SD RDT 206 51 57 429 80.2 (74.8–84.9) 88.3 (85.1–91.0)
4–8 159 11 24 93 93.5 (88.7–96.7) 79.5 (71.0–86.4)
9–14 69 1 2 23 98.6 (92.3–100.0) 92.0 (74.0–99.0)
15+ 232 15 73 495 93.9 (90.2–96.6) 87.1 (84.1–89.8)
0–14 434 63 83 545 87.3 (84.1–90.1) 86.8 (83.9–89.3)
0–3 IgM ELISA 83 166 41 540 33.3 (27.5–39.6) 92.9 (90.6–94.9)
4–8 143 28 19 124 83.6 (77.2–88.8) 86.7 (80.0–91.8)
9–14 68 2 1 24 97.1 (90.1–99.7) 96.0 (79.7–99.9)
15+ 199 48 41 530 80.6 (75.1–85.3) 92.8 (90.4–94.8)
4–14 211 30 20 148 87.6 (82.7–91.4) 88.1 (82.2–92.6)
0–3 IgG ELISA 40 216 45 530 15.6 (11.4–20.7) 92.2 (89.7–94.2)
4–8 79 90 18 127 46.7 (39.0–54.6) 87.6 (81.1–92.5)
9–14 56 14 2 25 80.0 (68.7–88.6) 92.6 (75.7–99.1)
15+ 226 24 56 508 90.4 (86.1–93.8) 90.1 (87.3–92.4)
4–14 135 104 20 152 56.5 (49.9–62.9) 88.4 (82.6–92.8)
a

The numbers of true positives (TP), false negatives (FN), false positives (FP), and true negatives (TN) of the test article when compared with the reference method are provided, and sensitivity [TP/(TP+FN)] and specificity [TN/(TN+FP)] are calculated. The upper and lower bounds for confidence intervals (95%) are also shown.