TABLE 2.
Comparison of IMMY GM ASR and MVista EIA for detection of Histoplasma antigen in urine using modified interpretive criteria (n = 150)
| IMMY GM ASR resulta | No. with MVista EIA result of: |
|
|---|---|---|
| Positive | Negative | |
| Positive | 14b | 0 |
| Negative | 3c | 121 |
| Indeterminated | 6 | 6 |
Positive agreement, 82.3% (95% CI, 58.2 to 94.6%); negative agreement, 100% (95% CI, 96.3 to 100%); overall agreement, 90.0% (95% CI, 84.6 to 94.0%); κ, 0.72 (95% CI, 0.55 to 0.89).
See Table S1 in the supplemental material for detailed clinical information.
See Table 4 for detailed clinical information.
For samples with indeterminate results by the IMMY GM ASR, the results of testing with the MVista EIA were applied; these results were not included for calculation of positive and negative agreement.