Table 1. Operating characteristics of flow cytometric methods for CD4+ T-cell enumeration, using conventional flow cytometers.

FLOW CYTOMETRY USING CONVENTIONAL FLOW CYTOMETER a
Conventional dual platform Bead-based single platform technology on conventional flow cytometer Panleucogating
Assay name Various Trucount tubes Flow-Count beads Perfect-Count PLG CD4
Manufacturer Various Becton Dickinson Beckman Coulter Cytognos Beckman Coulter
Compatible with various reagent systems and flow cytometers Compatible with various reagent systems and flow cytometers
Instrumentation Flow cytometer plus haematology analyser Flow cytometer Gating strategy
Assay principle Absolute count calculated using results from flow cytometry together with the total lymphocyte count from haematology analyser Absolute CD4 counts determined using ratio of CD4 to a known quantity of fluorescent beads—no need for haematology analyser Gating strategy using total white cell count
Trucount tubes contain premeasured quantity of lyophilised beads Liquid beads need accurate pipetting by operator Two different bead populations allow detection of inadequate mixing
Parameters measured Absolute CD4 and CD4% Absolute CD4 and CD4% Absolute CD4 and CD4%
Others depend on reagent kits/methods
Specimen 500μl whole blood in EDTA Up to 100μl whole blood in EDTA 100μl whole blood in EDTA
Throughput Up to 250 samples/day Up to 250 samples/day Up to 250 samples/day
Compatible with independent EQA/PT programmes? Yes Yes Yes Yes Yes
Is there access to compatible full process QC reagent? Yes Yes Yes Yes N/A
Maximum length of time from blood draw to testing 24 hours 24 hours 5 days
Can the test be performed on fixed/ stabilised blood Yes Yes Yes Yes Yes
Complexity/training required Complex. Significant training required. Complex. Significant training required Moderate training requirement
Environmental/energy issues Requires uninterrupted mains electricity Requires uninterrupted mains electricity Requires uninterrupted mains electricity
Robustness to heat/humidity Climate control recommended Climate control recommended Climate control recommended
Reagents require refrigeration Reagents require refrigeration Reagents require refrigeration

a Various flow cytometers available from various manufacturers. Examples include Becton Dickinson FACSCalibur, Beckman Coulter Epics XL.