Skip to main content
. Author manuscript; available in PMC: 2015 Mar 20.
Published in final edited form as: Nat Protoc. 2014 May 8;9(6):1301–1319. doi: 10.1038/nprot.2014.075

TABLE 2.

Troubleshooting table.

Step Problem Possible reason Solution
1 Labeling reaction is not successful or efficiency too low Protein samples may contain primary amine components that compete with labeling reagents Perform complete dialysis of protein sample to make sure protein samples are free of primary amines bearing reagents. Prepare fresh regents
2-4 The “heavy” to “light” ratio of the end time point mixture is not 1:1 Problem during labeling rection and mixing of “heavy” and “light” samples Ensure precision during pipetting, resuspension, mixing and transferring of samples
Problem of not performing “Experimental time-point” and “Reference” labeling reactions under identical condition Perform “Experimental time-point” and “Reference” labeling reactions side by side, and essentially, under identical conditions
Problem during mixing of “heavy” and “light” samples Be sure to equally mix experimental time-point and reference labeling samples
10 No signal, or poor signal intensity Possible protein sample loss during extraction and precipitation Be sure to mix and vortex as directed before centrifugation to reach equilibrium of extraction. Using gel-loading tips, carefully remove the aqueous layer without touching the finer emulsion that is created between the layers, containing protein of interest
21: A (x), B (xiv) and C (iv) No signal, or poor signal intensity Problem during desalting process Ensure that all buffers needed for desalting are appropriately prepared, as directed
Ensure that the pH of the reconstituted peptide is below 2 by pipetting 1 μL onto a pH strip; if not add more TFA
Ensure that air is not drawn into the column-tip and that it does not dry during the process
24 No signal, or poor signal intensity Sample-matrix crystallization process is not optimized or irregularities of spotting on MALDI plates Optimize sample-matrix crystallization conditions (See ref. Cohen et. al.)
Always use high-quality, ultra pure reagents and freshly prepared matrix
Take extra care when spotting and spot as uniform as possible
24 No signal, or poor signal intensity Possible problem of having low signal-to-noise of peptide ions on MS Increase laser to obtain optimized signal