Effect of AIG (200 μg/mL) on LTx quantification

Level Expected (ng/mL) No AIG AIG
LTx (ng/mL) SD %error LTx (ng/mL) SD %error
S1 24.3 22 5.20 9.0 17 0.79 32
S2 14.6 13 2.84 10 13 2.3 11
S3 9.7 11 1.34 16 8.5 2.8 12
S4 4.85 4.1 1.68 17 3.7 1.8 24
S5 2.43 2.6 0.108 8.1 1.6 0.68 35
S6 0.97 0.87 0.008 10 0.41 0.033 57
S7 0.485 0.48 0.014 1.9 0.20 0.014 58
S8 0.243 0.26 0.020 5.3 0.14 0.042 41
S9 0.146 0.16 0.030 11 0.096 0.029 34
S10 0.097 0.099 0.0035 2.5 0.042 0.012 57
S11 0.049 0.049 0.0060 0.03 0.016 0.0080 68
S12 0.024 0.018 0.0008 24 0.014 0.0071 42
S13 0.012 0.013 0.0017 9.0 <LOD NA NA
S14 0.0049 <LOD NA NA <LOD NA NA
S15 0.0024 <LOD NA NA <LOD NA NA
Plasma BL 0 <LOD NA NA <LOD NA NA
QCH 5.82 7.4 1.2 28 3.1 0.029 47
QCM 0.582 0.54 0.021 7.4 0.29 0.051 51
QCL 0.0582 0.06 0.001 0.21 0.02 0.010 70
Sample 88 88 2.6 0.23 93 4.7 5.7
Mean ± SD 9.4 ± 8.0 40 ± 20

Mean, SD, and accuracy (%error) for standards, QCs, and a clinical anthrax sample are shown