Table 3.
Cells | Label | Treatment | Notes | |
---|---|---|---|---|
Assay controlsa | Teff | CFSE | +acc, no stimulusd | Negative control: spontaneous Teff proliferation |
Teff | CFSE | +acc + stimulus | Positive control: maximum Teff proliferation | |
Treg | Claret | +acc, no stimulus | Negative control: spontaneous T regulatory proliferation | |
Treg | Claret | +acc + stimulus | Define Treg proliferation | |
Teff | CFSE | PHA | Positive control: verify that Teff are able to proliferate when cell surface receptors are bypassed | |
acc | None | None | Optional: confirm that they remain CD4− throughout assay and will not be confused with highly divided Teff | |
Teff | Varying CFSE | +acc + stimulus | Staining control: used during assay setup to verify that concentration of tracking dye chosen does not affect Teff proliferation (Fig. 1c)e | |
Treg | None | +acc + stimulus | Staining control: used during assay setup to verify that tracking dye labeled cells proliferate equivalently to unlabeled cells (Fig. 9)e | |
Instrument controlsb | Teff | None | Unstimulated, accessory only | Set voltage and negative region in CFSE channel |
Teff | CFSE | Unstimulated | Set voltage in CFSE channel; set color compensation in other channels; estimate location of undivided Teff | |
Treg | None | Unstimulated | Set voltage and negative region in Claret channel | |
Treg | Claret | Unstimulated | Set voltage in Claret channel; set color compensation in other channels; estimate location of undivided Treg | |
Teff (no Claret) | CD4 PE-Cy7 | Unstimulated | Set compensation | |
acc + Teff (no CFSE)c | LIVE/DEAD Fixable Violet | Unstimulated | Set compensation |
Assay controls are included in the experimental plate with test samples to verify that the expected biological outcomes can be recognized using the chosen instrument conditions
Instrument controls are used to establish instrument voltages and settings
Irradiated accessory cells will be nonviable and should be 100% positive for this viability probe
Acc = accessory cells (CD4 negative lymphocytes); stimulus = anti-CD3 plus anti-CD28
Needed only to establish optimized staining conditions for tracking dye when assay is first being implemented in the laboratory; not required on a routine basis