Table 1.

Results of a computational fluid dynamics simulation and pulse-wave ultrasonography measurements for patient 1 before and after virtual Pipeline embolization device deployment

Variable Pre-deployment
Post-deployment
Change (%)
CFD US-PSV (mean±SD) CFD US-PSV (mean±SD) CFD
Volumetric flow rate (mm3/sec)
  ICA 1,228.5 - 1,228.5 - -
  Aneurysm 598.2 - 59.9 - -90.0
  MCA 950.2 - 960.1 - 1.0
  PComA 278.3 - 268.4 - -3.6
Peak aneurysm pressure (mm Hg) 124.6 - 123.6 - -0.8
Peak aneurysm shear stress (N/m2) 18.0 - 14.5 - -19.4
Peak systolic velocity (cm/sec)
 Parent and outflow vessel
  ICA 38.2 37.3±2.6 39.0 37.9±2.4 2.1
  MCA 87.7 79.3±7.9 83.5 78.2±7.2 -4.8
  PComA 72.3 70.3±9.9 71.8 69.3±9.2 -0.7
 Aneurysm
  S1 -3.5 一4.3±1.2 1.3 1.8±0.8 -62.9
  S2 4.2 5.3±1.1 1.1 2.5±1.2 -73.8
  S3 -17.3 -16.5±2.1 -1.7 -3.0±0.8 -90.2
  S4 -23.6 -22.7±3.4 -3.4 -4.0±0.7 -85.6
Mean velocity (cm/sec)
 Parent outflow vessel
  ICA 9.0 10.8±2.1 9.4 10.5±1.9 4.4
  MCA 21.4 22.3±1.5 20.8 21.7±1.6 -2.8
  PComA 16.4 17.9±1.8 15.8 16.5±1.2 -3.7
 Aneurysm
  S1 -0.5 <1.0 0.2 <1.0 -60.0
  S2 0.8 <1.0 0.1 <1.0 -87.5
  S3 -3.4 -4.1±1.0 -0.4 <1.0 -88.2
  S4 -3.9 -4.5±1.2 -0.8 <1.0 -79.5

CFD, computational fluid dynamics; US-PSV, pulse-wave ultrasonography; ICA, internal carotid artery; MCA, middle cerebral artery; PComA, posterior communicating artery; S1-4, four quadrants of the aneurysmal sac.