Table 2.
Intraday accuracy and precision for analysis of 5-FU in blood and brain microdialysate.
| Nominal conc. (µg/mL) | Observed conc. (µg/mL) | Precision (%) | Accuracy (%) |
|---|---|---|---|
| Blood | |||
| 0.5 | 0.55 ± 0.09 | 16.72 | 9.76 |
| 1 | 1.05 ± 0.10 | 9.29 | 5.45 |
| 5 | 5.05 ± 0.31 | 6.19 | 1.05 |
| 10 | 10.16 ± 0.38 | 3.75 | 1.56 |
| 50 | 49.42 ± 0.84 | 1.70 | −1.16 |
| 100 | 100.27 ± 0.34 | 0.34 | 0.27 |
| Brain | |||
| 0.1 | 0.10 ± 0.01 | 11.45 | 4.94 |
| 0.25 | 0.24 ± 0.01 | 3.98 | −3.53 |
| 0.5 | 0.51 ± 0.02 | 4.55 | 1.44 |
| 1 | 1.06 ± 0.01 | 1.09 | 6.26 |
| 5 | 4.88 ± 0.06 | 1.21 | −2.30 |
| 10 | 10.05 ± 0.03 | 0.27 | 0.49 |
Data were expressed as means ± S.D. (n = 6).