Skip to main content
. 2015 Mar 10;2015:729679. doi: 10.1155/2015/729679

Table 2.

Intraday accuracy and precision for analysis of 5-FU in blood and brain microdialysate.

Nominal conc. (µg/mL) Observed conc. (µg/mL) Precision (%) Accuracy (%)
Blood
 0.5 0.55 ± 0.09 16.72 9.76
 1 1.05 ± 0.10 9.29 5.45
 5 5.05 ± 0.31 6.19 1.05
 10 10.16 ± 0.38 3.75 1.56
 50 49.42 ± 0.84 1.70 −1.16
 100 100.27 ± 0.34 0.34 0.27
Brain
 0.1 0.10 ± 0.01 11.45 4.94
 0.25 0.24 ± 0.01 3.98 −3.53
 0.5 0.51 ± 0.02 4.55 1.44
 1 1.06 ± 0.01 1.09 6.26
 5 4.88 ± 0.06 1.21 −2.30
 10 10.05 ± 0.03 0.27 0.49

Data were expressed as means ± S.D. (n = 6).