Table 3. Hysteresis and indentation depth values (M±SD) obtained at same ranges of applied loads before and after riboflavin/UV-A corneal cross-linking.
Hysteresis (pJ) | ||||||||||
Cornea 1 load range: 1.9±0.1 μN | Cornea 2 load range: 3.6±0.1 μN | Cornea 3 load range: 1.4±0.1 μN | Cornea 4 load range: 4.5±0.1 μN | Cornea 5 load range: 6.7±0.1 μN | ||||||
Approach speed (μm/s) | Before treatment | After treatment * | Before treatment | After treatment * | Before treatment | After treatment * | Before treatment | After treatment * | Before treatment | After treatment * |
1.7 | 0.29±0.05 | 0.17±0.03 | 0.35±0.07 | 0.30±0.04 | 0.26±0.02 | 0.21±0.01 | 1.26±0.03 | 0.82±0.01 | 2.40±0.69 | 1.88±0.02 |
3.5 | 0.35±0.02 | 0.21±0.02 | 0.49±0.03 | 0.36±0.06 | 0.30±0.03 | 0.25±0.01 | 1.53±0.02 | 0.99±0.02 | 2.36±0.04 | 1.78±0.03 |
8.2 | 0.42±0.03 | 0.28±0.04 | 0.66±0.03 | 0.53±0.02 | 1.47±0.02 | 1.16±0.01 | ||||
12.3 | 0.47±0.02 | 0.34±0.03 | 0.81±0.04 | 0.66±0.01 | ||||||
Indentation depth (μm) | ||||||||||
Cornea 1 load range: 1.9±0.1 μN | Cornea 2 load range: 3.6±0.1 μN | Cornea 3 load range: 1.4±0.1 μN | Cornea 4 load range: 4.5±0.1 μN | Cornea 5 load range: 6.7±0.1 μN | ||||||
Approach speed (μm/s) | Before treatment | After treatment | Before treatment | After treatment | Before treatment | After treatment | Before treatment | After treatment | Before treatment | After treatment |
1.7 | 2.40±0.04 | 1.97±0.20 | 2.46±0.04 | 2.18±0.08 | 2,03±0.63 | 1.33±0.04 | 2.75±0.04 | 2.17±0.05 | 2.12±0.28 | 1.97±0.04 |
3.5 | 2.21±0.05 | 1.85±0.09 | 2.35±0.02 | 2.03±0.04 | 1.45±0.30 | 1.28±0.11 | 2.50±0.04 | 2.01±0.02 | 1.60±0.02 | 1.38±0.05 |
8.2 | 2.02±0.03 | 1.72±0.13 | 2.24±0.03 | 1.89±0.03 | 2.10±0.01 | 1.97±0.04 | ||||
12.3 | 1.93±0.02 | 1.69±0.20 | 2.15±0.01 | 1.86±0.03 |
* Hysteresis values: P<0.05 between the pre- and post-treatment states