Skip to main content
. Author manuscript; available in PMC: 2015 Apr 3.
Published in final edited form as: Diagn Microbiol Infect Dis. 2014 Nov 27;81(2):102–104. doi: 10.1016/j.diagmicrobio.2014.11.010

Table 1.

Ocular swab pooling results on the Cepheid GeneXpert CT/NG Assay as compared to GenProbe AC2 for the detection of CT.

No. of pools with indicated result
Test No. Of pools tested R+/C+ R−/C− R+/C− R−/C+ % Sensitivity (95% CI) % Specificity (95% CI) NPV (%) PPV (%)
Gene Xpert 25 7 18 0 0 100 (56–100) 100 (78–100) 100 100

R = reference method result (AC2); C = comparative method result (GeneXpert); 95% CI = 95% confidence intervals.