Table III.

Concordance parameters of the three POCT methods in relation to the reference method.

DiaSpect NBM-200 HemoControl
Bias (g/dL) (95% CI) −0.18 (−0.27 to −0.09) −0.66 (−0.78 to −0.53) −0.22 (−0.31 to −0.13)
Upper limit of agreement (g/dL) (95% CI) 1.68 (1.54 to 1.83) 2.09 (1.88 to 2.30) 1.7 (1.55 to 1.85)
Lower limit of agreement (g/dL) (95% CI) −2.05 (−2.19 to −1.91) −3.39 (−3.61 to −3.19) −2.15 (−2.30 to −1.99)
ICC coefficient (g/dL) (95% CI) 0.78 (0.74 to 0.82) 0.43 (0.27 to 0.55) 0.77 (0.73 to 0.81)
Coefficient of variation (%) 2.19 4.28 2.51

POCT: point-of care testing; CI: confidence interval.