Table 3.

Percentage-error and bias CItd-CIpc values after 1 h, 2 h, 4 h, 6 h and 8 h without thermodilution (one data set per patient)

Evaluation Validation Merged data
Interval to last thermodilution Percentage Error Mean bias [L/min/m2] Percentage Error Mean bias [L/min/m2] Percentage Error Mean bias [L/min/m2]
1 h 23.2% −0,144 ± 0,501 23.4% 0.026 ± 0.465 23.5 −0.039 ± 0.483
2 h 18.6% −0,021 ± 0,400 32.5% 0.112 ± 0.660 27.7 0.061 ± 0.576
4 h 28.5% −0,020 ± 0,618 24.4% 0.097 ± 0.490 26.1 0.054 ± 0.539
6 h 30.9% −0,248 ± 0,690 40.3% −0.200 ± 0.861 36.6 −0.218 ± 0.794
8 h 27.9% 0,082 ± 0,600 31.0 0.033 ± 0.632 29.6 0.062 ± 0.615