Table 3. The BD GeneOhm Cdiff assay and the multiplex real-time PCR assay results compared to toxigenic culture as the reference method.

Assay Toxigenic culture Assay performance (95% confidence interval)
Negative Positive Sensitivity (%) Specificity (%) PPV (%) NPV (%)
BD GeneOhm Cdiff
 Negative 298 1 97.8 (87.0-99.8) 97.7 (95.1-98.9) 86.5 (73.5-93.9) 99.6 (97.8-99.9)
 Positive 7 45
 Total 305 46
Real-time PCR
 Negative 301 2 95.6 (83.9-99.2) 98.6 (96.4-99.5) 91.6 (79.1-97.2) 99.3 (97.3-99.8)
 Positive 4 44
 Total 305 46
P value* 0.361 > 0.999 > 0.999 0.413

*P value for the comparison of sensitivity, specificity, PPV, and NPV of BD GeneOhm Cdiff and Real-time PCR assays.

Abbreviations: PPV, positive predictive value; NPV, negative predictive value.