Published in final edited form as: Heart Lung. 2014 Dec 24;44(2):129–136. doi: 10.1016/j.hrtlng.2014.07.007

Table 5.

Sensitivity versus specificity in a 30-day evaluation framework for ≥ 1, ≥2 and ≥3 observations for the parameter set excluding OptiVol.

Number of device observation(s) Sensitivity
Specificity
Unadjusted GEE adjusted (95% CI) Unadjusted GEE adjusted (95% CI)
≥ 1 observation(s) 28/42 (66.7%) 68.9% (52.8-81.5) 1600/2234 (71.6%) 71.2% (68.4-73.9)
≥ 2 observations 11/42 (26.2%) 27.0% (15.2-43.3) 2118/2234 (94.8%) 94.5% (93.1-95.7)
≥ 3 observations 4/42 (9.5%) 9.5% (3.7-22.5) 2209/2234 (98.9%) 98.8% (98.2-99.3)