Problems caused during imaging |
|
“Pump” or “Stokes” beam wavelength used is incorrect
Light is not focused on the sample
AOM is not working
Photodiode is improperly functioning
|
Re-calculate difference frequency. Check that the lasers are tuned to vibrational resonance.
Calibrate with pure reference chemical (dodecane/triolein) prior to imaging samples.
Scan a range of wavelengths (± 2 nm).
Ensure proper mirror alignment and condenser placement (critical).
Ensure that AOM is functioning (check with a fast photodiode). Re-check settings of lock-in amplifier, reference to right modulation frequency.
Removing excitation filter; photodiode should obtain a strong signal from the modulated excitation beam.
|
|
Laser is not aligned to the center of the scanning area
Laser power is too low/too high
Amplifier settings are incorrect
Low/high lipid levels in the sample
|
The laser should be aligned to the center of the scanning area using dodecane oil and by adjusting the mirror angles, before imaging samples. If changing the laser power or amplifier settings cannot increase the signal, then calibrate the setup again.
On the image acquisition software, set the pixel intensity such that the background will be almost zero. Also avoid excess saturated pixels.
A general rule of thumb is to image wild-type/control samples first, arrange the settings accordingly, and keep those parameters constant during the analysis of mutant/treated samples.
|
|
Amplifier/detector settings are incorrect
Transferred excess bacteria when picking worms
Cell culture medium generates background signal
|
Minimize transfer of bacteria while picking the worms, as this may contribute to high background.
Replace with alternate medium/PBS. Image as soon as possible.
|
|
|
|
Problems caused during sample preparation |
|
|
Wait for several minutes after adding worms to the anesthetizing solution.
Increase the concentration of sodium azide or levamisole modestly. If worms must be recovered after imaging, avoid highly concentrated solutions as they may kill worms.
|
|
The samples waited too long for imaging after preparation
Volume of the anaesthetizing solution is too low or its concentration too high
Concentration of the agarose solution used for the pad is too high
|
Whole worm samples need to be imaged within approximately 30 minutes. Preservation at 4°C for a few more hours longer is possible, but samples cannot be recovered.
If worms burst even within 30 minutes, the agarose pad or the anaesthetizing solution might be causing the problem. Lower the concentration of the agarose or the anaesthetizing solution. Also try increasing the volume of anesthetizing solution.
|
|
|
|